| Environmental Temperature and Humidity Testing Chamber | ||||||
| Secifications ( at +20°C room temp. Or+25°C water temp with empty load) | ||||||
| Model | KMH-L series | |||||
| 150L | 225L | 408L | 800L | 1000L | ||
| Interior sizel | W(mm) | 600 | 700 | 700 | 1000 | 1000 |
| H(mm) | 600 | 700 | 750 | 1000 | 1000 | |
| D(mm) | 460 | 480 | 800 | 800 | 1000 | |
| Exterior size | W(mm) | 880 | 980 | 980 | 1280 | 1280 |
| H(mm) | 1730 | 1890 | 1940 | 2160 | 2160 | |
| D(mm) | 1400 | 1460 | 1740 | 1750 | 1950 | |
| Weight(KG) | 250 | 300 | 420 | 530 | 600 | |
| Temp. range | -40 °C ~ +150°C | |||||
| Humidity Range | 20 ~ 98&R.H. | |||||
| Temp. constancy | ±0.5ºC | |||||
| Humidity constancy | ±2.5%R.H | |||||
| Humidity uniformity | ±3.0 % R.H | ±5.0 % R.H | ||||
| Heat up time | -40°C ~+100°C , within 45 min | |||||
| Pull down time | +20°C ~ -40°C , within 60 min | |||||
| Equipment Power | 7.5 | 7.5 | 8.5 | 14.5 | 14.5 | |
| Ambient Temp. | 5ºC ~+35ºC | |||||
| Standard accessories | Sight window*1, cable port(portΦ50mm)*1, shelves*2, chamber lamp*1, power cable *2m | |||||
| Safety device (standard) | No fuse breaker, over pressure \ over heat and over current protection for compressor, Over temp. protect, over load protect of blower, dry heat protection | |||||
| Control System | Balanced Temperature & Humidity Control System | |||||
| Refrigeration system | Air Cooled ( Water Cooled Type is option ) | |||||
| Hermetic compressor,Single stage refrigeration system, CFC-Free refrigerant(HFC-507/HFC-23) | ||||||
| Controller | Panel:7-inch LCD Touch panel, Chinese or English display selectable | |||||
| Operation model :Program or fix point running | ||||||
| Memory capacity :120programs,1200steps, | ||||||
| All repeat 999 cycles, part repeat 99 cycles | ||||||
| Interior material | Stainless Steel Plate ( SUS 304 ) | |||||
| Exterior material | Baked Painting Steel Or Stainless steel(SUS304) | |||||
| Insulation material | Rigid polyurethane Foam and Glass Fiber Wool | |||||
| System | Balanced & Humidity Control System | |||||
| Wiring method | AC 380±10 % 50Hz ,3 phase 4 wires +Ground Wires | |||||
| Note: 1. We reserve the right to change specifications without prior notice 2. Customized sizes and configurations available | ||||||